Update on recent advances in technology in type 1 diabetes

AISHA ASLAM,1 HOOD THABIT,1,2 LALANTHA LEELARATHNA1,2

1 Diabetes, Endocrinology and Metabolism Centre, Manchester Royal Infirmary, Manchester University NHS Foundation Trust & Manchester Academic Health Science Centre, Manchester, UK
2 Division of Diabetes, Endocrinology and Gastroenterology, Faculty of Biology, Medicine and Health, University of Manchester, UK

Address for correspondence: Dr Lalantha Leelarathna
Diabetes, Endocrinology and Metabolism Centre, Manchester Royal Infirmary, Manchester University NHS Foundation Trust Manchester, Oxford Rd, M13 9WL, UK.
E-mail: Lalantha.leelarathna@mft.nhs.uk

Br J Diabetes 2024;24(1):30-37

https://doi.org/10.15277/bjd.2024.440

Abstract

Those living with type 1 diabetes (T1DM) require daily adjustments of exogenous insulin doses and frequent glucose monitoring to optimally manage their condition. Consequently, it is one of the most challenging long-term conditions to live with. Recent years have seen major progress in the management of T1DM, with minimally invasive glucose monitoring technology and glucose-responsive insulin delivery systems, also called hybrid closed-loop systems. This narrative review focuses on three key areas: continuous glucose monitoring (CGM), hybrid closed-loop (HCL) systems, and connected pen devices, sometimes known as smart pens. We describe features of commonly used devices in the UK NHS and summarise their key evidence base. Randomised controlled trials and real-world studies of CGM devices have shown improved haemoglobin A1c (HbA1c) levels, improved sensor-based metrics such as higher time spent in the target glucose range, and reduced rates of hypoglycemia. HCL studies have similarly shown improved HbA1c and other sensor- based glucose outcomes. Further recent innovations for insulin users include connected insulin pens, which allow the display and recording of insulin delivery information. In addition to glycaemic benefits, novel diabetes technology has been shown to improve quality of life and to give higher treatment satisfaction. Some disadvantages of technology include alarm burden, connectivity problems and premature device failure. To get the best from novel diabetes technology, appropriate training and education are required, specifically in identifying and dealing with critical system failures such as cannula failure and the risk of ketoacidosis. Recent recommendations from the National Institute of Health and Care Excellence (NICE) regarding HCL further underscore the growing significance of these advances in diabetes care.

Key words: diabetes technology, continuous glucose monitoring, hybrid closed loop

Introduction

Type 1 diabetes (T1DM) is a lifelong endocrine condition related to insulin deficiency caused by pancreatic beta-cell dysfunction.1 People with T1DM need to be able to self-adjust their insulin dose to achieve normoglycaemia and minimise the risk of micro- and macrovascular complications. Glucose monitoring entails multiple glucose measurements per day to facilitate self-adjustment of insulin and is thus a key foundation of modern T1DM self- management.2 The pain and inconvenience of conventional glucose monitoring through frequent finger-stick capillary blood testing however is, unsurprisingly, associated with lower quality of life, and poor treatment satisfaction and adherence.3,4

Advances in glucose monitoring

Capillary blood glucose meters were introduced into clinical practice in the late 1960s. For years, diabetes management consisted primarily of daily fingerstick testing, multiple daily insulin administrations, and handwritten diaries to record glucose levels. Current sensor-based glucose monitoring devices, first introduced in the 1990s, provide a minimally invasive method to measure interstitial fluid glucose levels in real time. Commercial externally worn devices use a subcutaneously implanted needle-type amperometric enzyme electrode: this measures interstitial glucose concentration by detecting changes in the electric current caused by the enzyme-catalysed oxidation of glucose into hydrogen peroxide.5

Since the introduction of the first continuous glucose monitoring (CGM) system by Medtronic in 1999, innovations in CGM have undergone a transformative change. Flash glucose monitoring, more specifically the Freestyle Libre System (FSL), was first introduced in 2014 in Europe. It consists of a 2-week, externally worn glucose sensor that displays present, 8-hour historic and trend glucose data when the user “scans” the sensor using nearfield communication via a Freestyle Libre reader or compatible mobile phone.6,7 The first generation of FSL differs from real-time continuous glucose monitoring systems (RT-CGM) in that FSL does not continuously display or transmit sensor glucose information unless physically scanned by the user and does not provide low or high glucose alarms. The second generation FSL (Freestyle Libre 2), first released in Germany in 2018, incorporates a Bluetooth transmitter in addition to near-field communication, continuously sending data to activate low and high glucose threshold alerts on the Freestyle Libre 2 reader without needing to scan. However, the user is still required to scan to visualise the actual glucose value, causing the alerts. In the UK, the Freestyle Libre 2 app was upgraded in August 2023 to display glucose data without having to scan the sensor. (Scanning is still required with the Libre 2 reader).8

The assessment of CGM accuracy is not standardised, which poses challenges for regulators, clinicians and users. The mean absolute relative difference (MARD) is the most commonly used measure. MARD is based on the comparison between paired measurements of a given CGM system and a reference method.9 As measured by MARD, the accuracy of CGM devices has improved over time, reducing from 25% in 1990 to <10% in recent years. Kovatchev and colleagues, using simulation techniques, have calculated a minimal accuracy of a mean ARD of <10% for real-time CGM to reach sufficient safety when sensor glucose data are used for insulin dosing decisions.10 Although MARD is widely used, there are several limitations to MARD. It is affected by study design, level of hypoglycemia, hyperglycaemia and glucose variability. Therefore, comparing MARD between studies should be done with caution. Additional sensor accuracy measures such as Clarke or Consensus error grids and precision Absolute Relative Difference (PARD) where two similar CGM systems are used as a reference, can be useful tools to assess CGM accuracy.11

Table 1 (adapted from https://www.diabetesspecialistnurseforumuk.co.uk/whats-new CGM Comparison chart from Diabetes Specialist Nurse Forum UK) summarises the key features of various CGM devices commonly used in the UK NHS.12

1191 Aslam Table 1

Evidence from clinical trials

Real-time CGM studies
The landmark JDRF multicentre randomised controlled trial (RCT) examined the benefits of using CGM using earlier generations of sensors (Dexcom Seven (Dexcom), the Minimed Paradigm System (Medtronic) or the Freestyle Navigator (Abbott Diabetes Care). In this study involving 322 people with T1DM with HbA1c levels above 7.0% (53 mmol/mol), CGM led to improved HbA1c levels for individuals over 25 years old. There was no improvement noted for those under 25, and this was linked with lower sensor use in this age group. In a secondary study, which included 129 people with HbA1c levels below 7.0% (53 mmol/mol), CGM helped individuals maintain their target HbA1c levels while reducing exposure to hypoglycaemia.13

The GOLD study was an open-label crossover RCT in 161 people with T1DM on multiple daily injections (MDI). The study compared the utility of Dexcom G4 CGM (Dexcom Inc. San Diego, US) with self-monitoring blood glucose (SMBG). The mean HbA1c was 7.9% (63 mmol/mol) with CGM vs 8.3% (67 mmol/mol) with SMBG. The mean difference of −0.43% (4 mmol/mol) is thought to be a clinically meaningful reduction. There was only one severe hypoglycemia incident in the CGM group, compared to five such incidents in the SMBG group.14

The Diamond study was a parallel design multicentre trial in which 2:1 randomisation was used for 158 people with T1DM on MDI, for CGM Dexcom G4 (Dexcom Inc., San Diego, US) and conventional therapy. The improvement in HbA1c was 1.1% (11.6 mmol/mol) at 12 weeks and 1.0% (10.6 mmol/mol) at 24 weeks with CGM vs 0.5% (5.3 mmol/mol) and 0.4% (4 mmol/mol) with conventional therapy. A mean difference of 0.6% (6.5 mmol/mol) was noted between the two groups at 24 weeks. Although both groups had similar incidences of severe hypoglycaemia incidents, time spent in the hypoglycaemia range was lower in the CGM group.15

Another RCT in 203 older adults (average age >60 years) with T1DM showed that median time with glucose <4 mmol/L (72 mg/dl) was reduced from 5.1% at baseline to 2.7% at six- month follow-up in the CGM group (difference of 1.9% between the two groups at six months), demonstrating the promise of CGM in reducing hypoglycaemia burden in this age group.16

Further, a parallel design RCT was conducted in 153 young adults (mean age 17 years) with T1DM. The study compared Dexcom 5 (Dexcom Inc. San Diego, US) CGM with SMBG over 26 weeks. The study showed a reduction in HbA1c level from 8.9% (74 mmol/mol) to 8.5% (69 mmol/mol) in the CGM group, compared to no difference in HbA1c level in the control group.17 In a crossover RCT conducted in adolescents and young adults, Dexcom G6 CGM (Dexcom Inc. San Diego, US) was compared with SMBG. The study showed that eight weeks of CGM use improved time in range (TIR) by 11.1% (CGM 35.7% vs. SMBG 24.6%), reduced mean sensor glucose by 32.2 mg/dL (1.8 mmol/L) [219.7 mg/dL (12.2 mmol/L) with CGM, 251.9 mg/dL (14.0 mmol/L) with SMBG], and lowered HbA1c levels by 0.76% (8 mmol/mol) compared to SMBG.18

CONCEPTT is a multicentre RCT in which women with T1DM who were either pregnant or planning pregnancy are assigned to RT-CGM and SMBG groups. In pregnant women, RT-CGM use showed improvement in HbA1c levels by 0.19% (2 mmol/mol), an increase in TIR (68% RT-CGM vs 61% SMBG), and less time in hyperglycemia (27% RT-CGM vs 32% SMBG group). Neonatal health outcomes also improved in the CGM group. However, RT- CGM did not show the same benefits in women planning pregnancy.19

Intermittently scanned CGM studies
The IMPACT trial was a multicentre RCT including 239 people with well controlled (HbA1c <7.5% [59 mmol/mol]) T1DM and intact awareness of hypoglycaemia. Intermittently scanned CGM (isCGM device) Freestyle Libre (Gen 1) was compared with SMBG. IsCGM use was associated with a 46% reduction in time spent in hypoglycaemia (<3.9 mmol/L or 70 mg/dl) at six months, without any change in total daily insulin dose. Treatment satisfaction was noted in the intervention group as compared to the control group.20

The FLASH-UK trial, a multicentre RCT of 156 people with T1DM, investigated the use of isCGM as an alternative to finger- prick testing in those with HbA1c >7.5% (59 mmol/mol). The isCGM group experienced a reduction in baseline HbA1c from 8.7% (72 mmol/mol to 7.9% (63 mmol/mol) at 24 weeks, with a 0.5% (5.3 mmol/mol) point difference between the two groups. The health-economic analysis of FLASH-UK revealed a small and statistically insignificant increase in Quality-Adjusted Life Years (QALY) during the trial, along with a short-term increase in costs. However, over the lifetime of a patient isCGM was projected to be cost-effective, particularly in those with high baseline HbA1c, with an aggregate cost per QALY (ICER) of £4,445, which is within the NICE-recommended cost-effectiveness threshold of less than £20,000 per QALY.21,22 ALERT1 was a RCT in people with T1DM (already on isCGM), which compared RT-CGM with isCGM. It showed a reduction in self-reported hypoglycaemia worry and improvements in TIR (59.6% RT-CGM vs 51.9% isCGM) and HbA1c (7.1% [54 mmol/mol] RT-CGM vs 7.4% [57 mmol/mol] isCGM) for the RT-CGM group. Though the investigators recommended the use of RT-CGM in the management of patients with T1DM, the study was not able to answer the question of whether those naive to CGM should start either RT-CGM or isCGM.23

Evidence from real-world data

Findings from real-world data have been consistent with RCTs, with improvements in glycaemic control, user satisfaction and hypoglycaemia awareness. The Association of British Clinical Diabetologists (ABCD) conducted a multicentre service evaluation for over two years. It showed that, irrespective of duration of use, improvement in HbA1c was evident (ranging from 0.37% [4 mmol/mol] to 0.55% [5.8 mmol/mol] improvement) in FSL users. There was further improvement in the GOLD score (hypoglycaemia awareness score) and DDS Score (reduction in diabetes distress) after FSL use.24

Another real-world observational study in 515 adults who used RT-CGM for at least two months showed improvements from baseline HbA1c and hospitalisation from ketoacidosis, and a reduction in hypoglycaemia from 16% to 4%. Admission days, work absenteeism and quality of life improved significantly, with a reduction in the fear of hypoglycaemia with RT-CGM use.25

Collectively, these studies highlight the utility of CGM in improving HbA1c, the time spent in the 'target glucose range' and the reduction in hypoglycaemia burden.

Advances in automated insulin delivery systems (hybrid closed loop)

Automated insulin delivery (AID) systems integrate insulin pumps with CGM devices using mathematical algorithms, allowing for glucose-responsive insulin delivery.26 At present, the system continues to be classified as a hybrid system, as it requires user input of mealtime carbs and delivery of a meal bolus.19

Key features of these systems are shown in Table 2 (Adapted from https://www.diabetesspecialistnurseforumuk.co.uk/whats-new HCL comparison chart from Diabetes Specialist Nurse Forum UK).27

1191 Aslam Table 2

The commonly used AID systems in the UK NHS include:

  1. Minimed 670G/780G (Medtronic) (Medtronic 670 G with Guardian 3 and Medtronic 780G with Guardian 4 sensor). Manufacturer: Medtronic; country of origin US
  2. Tandem t:slim X2 Control IQ (T slim X2 pump with Dexcom G6 sensor). Manufacturer: Tandem Diabetes Care; country of origin: US
  3. CamAPS FX system (DANA and Ypsomed pump with DexcomG6/Libre 3). Manufacturer: CamDiab Ltd (Cambridge, UK), and Ypsomed (Ypsomed AG Switzerland) and DANA (Advanced Therapeutics UK) (SOOIL, South Korea). Country of origin: UK, Switzerland and South Korea
  4. Omnipod 5 (INSULET) (Omnipod 5 pump with Dexcom G6 and Libre 2 plus ). Manufacturer: Insulet Corporation; country of origin: US. At present, Omnipod 5 with Libre 2 Plus is not available in the UK. However, it is anticipated that it may become available in the latter half of 2024.

In addition to commercially available AID systems, a category of open-source AID systems has gained global acceptance, embraced by a substantial user base. These technologies are the result of user-initiated advancements facilitated by a committed online community affected by diabetes. Real-world evidence consistently validates the safety and efficacy of open-source AID systems, highlighting notable improvements in glycaemic control and positive impacts on aspects such as quality of life, sleep quality and fear of hypoglycaemia.28

The CREATE trial, a recent RCT, has further confirmed the benefit of open-source systems when compared with sensor- augmented pumps. The average TIR has improved by 3 hours and 22 minutes in the open-source AID system.29

Evidence behind commercial systems

CamAPS Fx
The CamAPS Fx algorithm is available with two different pumps: DANA and Ypsomed insulin pump. When used with the DANA pump, only the Dexcom G6 sensor is compatible, while with the Ypsomed pump users have the option of choosing between Dexcom G6 or Libre 3 sensor. The University of Cambridge group in the UK has performed a large number of randomised clinical studies over the last 10 years, including studies in children and adolescents,30 pregnant women,31,32 and adults.33

In 2018, an RCT over a 12-week period involving 114 people with T1DM (both children and adults) found that HCL therapy with the CamAPS System increased the proportion of time spent within the target range (mean difference 10.8% points between the two groups), decreased HbA1c levels (mean difference 0.31% or 3.3 mmol/mol), and shortened the duration of both low and high glucose concentrations.33

The AiDAPT Study, an RCT in pregnant women, found a higher percentage of time in the pregnancy-specific target glucose range (3.5-7.8 mmol/L or 63 mg/dl-140 mg/dl) (68.2% CamAPS group vs. 55.6% control group) and improvements in secondary outcomes, such as less time spent in hyperglycaemic states, more overnight time in the target range and lower HbA1c levels, among the closed-loop group of 124 pregnant women.32

Medtronic 780G
The FLAIR study, a 12-week crossover trial with 113 people with T1DM aged 14-29, compared the Medtronic 670G and advanced hybrid closed-loop (AHCL) systems. The 670G group showed a 6.0% TIR improvement compared to baseline, while the AHCL group exhibited a more significant 10% increase, indicating the superior impact of AHCL on glycaemic control.34

The ADAPT study was another RCT for patients with T1DM that compared AHCL with the combination of MDI and CGM. Mean HbA1c levels were reduced by 1.54% (SD 0.73) in the AHCL group, from 9.0% (75 mmol/mol) to 7.3% (56 mmol/mol). In comparison, the mean HbA1c levels were only reduced by 0.20% (SD 0.80) in the group receiving MDI plus isCGM, from 9.0% (75 mmol/mol) to 8.9% (74 mmol/mol). The model-based difference between the two groups was -1.42% (95% CI -1.74 to -1.10; p<0.0001).35

Tandem Control IQ
In 2019, an RCT was conducted on 168 individuals with T1DM. Among them, 112 were assigned to the closed-loop group (Control IQ system) and 56 to the control group (sensor-augmented pump). The participants' ages ranged from 14 to 71 years, and their initial HbA1c levels varied from 5.4% (36 mmol/mol) to 10.6% (92 mmol/mol). Over a span of six months, the closed-loop group improved their time within the target glucose range by 11 percentage points compared to the control group.36

Additional real-world data from Breton and colleagues confirmed results consistent with the RCT, with a 10-12% improvement in TIR using the Control IQ system.37

Omnipod 5 system
Omnipod 5 (Insulet) is a tubeless HCL system recently approved for use in the UK. No RCT of this system has been published. However, a single-arm before and after study suggested the safety and effectiveness of the system. Following a conventional treatment phase, 112 children and 129 adults used this system, resulting in an improvement in HbA1c levels (0.71% or 7.8 mmol/mol in children, 0.38% or 4.2 mmol/mol in adults) and an improvement in TIR by 15.6 ± 11.5% in children and 9.3 ± 11.8% in adults. Overall, the device was well tolerated and it improved glucose management.38

Evidence for HCL from the NHS

A real-world observational study, led by ABCD DTN (Diabetes Technology Network UK) and funded by NHS England (NHS England closed-loop pilot),39 investigated the use of the HCL system in a cohort of 570 patients with diabetes with HbA1c >8.5% (69 mmol/mol). The findings demonstrated an improvement in HbA1c levels by 1.7% (18 mmol/mol, p<0.0001), along with an increase in TIR from 34.2% to 61.9%. The proportion of individuals with HbA1c levels of 7.5% (58 mmol/mol) or lower rose from 0% to 39.4% (p < 0.0001). 94.7% of the participants reported a positive impact on their quality of life, indicating the potential of HCL systems to transform diabetes management.40

Connected pen technology
In the realm of diabetes management, connected pens and button devices have emerged as innovative tools that cater to patients requiring MDI. These devices employ Bluetooth, or near-field communication technology, to record insulin dosages and the timing of injections, offering convenience and improved data tracking for individuals with diabetes.41

An observational study from 2019 examined the use of a Bluetooth-enabled pen device by 75 people with MDI. The study revealed that a significant proportion of participants missed their bolus and basal insulin doses when using these devices, and this was found to be strongly linked to suboptimal glycaemic control. These findings highlight the significance of monitoring and resolution of adherence concerns in the management of diabetes to enhance overall glycaemic control.42

The usefulness of one connected pen device (Insulclock connected cap device) was shown by a RCT of 55 people with T1DM (26 active on the device vs. 29 masked). Those in the active group experienced a 5.2% increase in TIR vs. -0.8% in the masked group. The active group further showed lower average glucose levels, less time above the target range, and a higher on-time insulin dosage rate.43

This connected pen technology holds immense potential for simplifying insulin administration and data tracking, benefiting individuals with diabetes who rely on MDI. Nevertheless, the research emphasises the need for continuing investigation and user enlightenment to maximise the advantages offered by these ground-breaking instruments.

Non-glycaemic benefits, patient-reported outcomes and disadvantages of novel diabetes technology
Many studies evaluating diabetes technology also include the assessment of diabetes-related patient-reported outcomes (PROMs), with a focus on outcomes related to treatment satisfaction, hypoglycaemia burden and diabetes distress. However, findings from these studies have shown inconsistencies and depend on factors such as study type and the characteristics of the study population.44

Diabetes technology offers several benefits, such as glycaemic control, helping remove the burden of disease from the user and providing comprehensive data for healthcare providers to make more informed treatment decisions. It also comes with disadvantages. Commonly observed drawbacks include information overload, alarm fatigue (users being frequently bothered by alarms, whether real or false), the constant presence of sensors on the body, and the potential for skin irritation.45

Our understanding of the impact of HCL therapy on retinopathy outcomes remains limited. A 2021 NHS pilot study examined the effects of HCL on retinopathy outcomes in individuals with T1DM and an HbA1c >8.5% (69 mmol/mol). The study, encompassing a nine-month follow-up of 62 participants, identified a 12.5% retinopathy progression rate, highlighting the imperative need for continuous vigilance to discern and address potential risks.46

Users require better education to fully understand and utilise these systems. In cases of system failure, patients should be aware of sick day rules, monitor blood ketones, and manage blood glucose levels while administering insulin doses.

A significant challenge to the sustainable use of AID systems is ensuring user acceptance and helping individuals to integrate these technologies into their daily lives, addressing the various challenges that come with long-term use.19

NICE guidance

In 2022, the National Institute for Health and Care Excellence (NICE) updated its guidance on glucose monitoring for people with T1DM. They recommended providing either flash or continuous glucose monitoring to all people with T1DM.47

The NICE technology appraisal (TA 943) published in December 2023 advocates HCL systems as an efficacious method for regulating blood glucose levels. This recommendation extends to children, young people, and women who are either attempting conception or are pregnant. Furthermore, it is recommended for adults with T1DM exhibiting an HbA1c of 58 mmol/mol (7.5%) or higher, or those encountering severe hypoglycemia despite optimal management with continuous subcutaneous insulin infusion (CSII), RT-CGM or isCGM. These recommendations signify a substantial progression in the treatment of diabetes, heralding a notable improvement in managing the condition.48

Despite the publication of the NICE TA around HCL therapy, significant challenges exist in terms of implementation. In England, only 15% of people living with T1DM are using an insulin pump. It is not clear how each area will prioritise the implementation and it is likely that significant expansion of the workforce may also be required for full implementation of the guidance.

There is currently no dedicated NICE guidance on the use of connected pen devices, necessitating a future review of the evidence for efficacy and cost-effectiveness in the NHS. Pen devices have a potential role, particularly for those who may not be suitable candidates for conventional insulin pumps or HCL.

Evolution of health care professional role in response to recent technological innovations

As the use of technology increases in T1DM, health care professionals (HCPs) and educators will need to understand the role and suitability of these technologies in addressing the different clinical and personal needs of individuals, whilst discussing the benefits and limitations to ensure that appropriate expectations are met. HCPs providing AID therapy will also need to ensure that specific education and clinical guidance on managing glycaemia during periods of illness or exercise are provided to users of this system, to ensure their safe and effective use. Ultimately, HCPs can be advocates on behalf of people living with diabetes, to ensure wider and equitable access to technology, especially in underserved populations.

Future landscape for diabetes technology

Further progress to improve the performance and usability of closed-loop (CL) systems is underway. Examples of these are highlighted below:

Conventional single-hormone or insulin-only CL systems still carry a residual risk of hypoglycaemia due to the limitations of subcutaneous insulin action. Dual-hormone CL using glucagon as an adjuvant has the potential to further reduce this risk. An example of a dual-hormone system awaiting full regulatory approval is the iLET CL system, also known as the “Bionic Pancreas”.

Ongoing trials of the Bihormonal Pancreas AID system by Beta Bionics,49,50 promise a more comprehensive approach to blood glucose regulation and enhanced responsiveness to mitigate hypoglycaemic events.

1191 Aslam Key Messages

CGM usability and performance are other factors directly impacting CL users’ experience and outcome. Currently, minimally invasive subcutaneous CGM devices are key components of CL systems, with acceptable accuracy and wearability. However, research into the development of novel CGM sensors that may enhance CL use and reduce device burden is ongoing. An implantable CGM sensor, Senseonics (Senseonics Holdings Inc, NYSE America) has been approved for use as a standalone CGM device, but it is no longer available in the UK at present. Research into non-invasive CGM sensors which uses technology such as radio-frequency identification (RFID) shows promise.51 However, further clinical studies are needed to demonstrate acceptable accuracy and safety.

Access to greater interoperability and choice of different CL components (i.e. the ability to use different CGM sensors and insulin within a CL system) is another important factor in enhancing the usability and acceptance of CL therapy. The CamAPS FX CL system adopts a modular approach, allowing users to have a choice of more than one CGM and insulin pump device which are compatible for use with the algorithm. This approach underscores the potential for further advancements in CL use, enhancing customization and user choice.

Conclusion

There have been tremendous advances in diabetes technology in the last couple of decades, with novel glucose sensors and automated insulin delivery showing increased accuracy, convenience, improved glycaemic control and better quality of life for individuals with diabetes.

To unlock the full potential of these technologies for enhanced diabetes management, continuous research, effective user education and treatment tailored to each patient's specific needs are required. With these cutting-edge tools at our disposal, the future of diabetes treatment holds significant promise.

© 2024. This work is openly licensed via CC BY 4.0

© 2024. This work is openly licensed via CC BY 4.0.

This license enables reusers to distribute, remix, adapt, and build upon the material in any medium or format, so long as attribution is given to the creator. The license allows for commercial use. CC BY includes the following elements: BY – credit must be given to the creator.

Copyright ownership The author(s) retain copyright.

Conflict of interest AA has no conflicts of interest. HT receives consulting fees and speaker honoraria from Eli Lilly, and reports having received research support from Dexcom Inc. LL has received personal fees from Abbott Diabetes Care, Dexcom, Insulet, Medtronic and Novo Nordisk.

Funding None.

References

  1. Roep BO, Thomaidou S, van Tienhoven R, Zaldumbide A. Type 1 diabetes mellitus as a disease of the β-cell (do not blame the immune system?). Nat Rev Endocrinol 2021;17(3):150–61. https://doi.org/10.1038/s41574-020-00443-4
  2. McCarthy MM, Grey M. Type 1 Diabetes self-management from emerging adulthood through older adulthood. Diabetes Care 2018; 41(8):1608–14. https://doi.org/10.2337/dc17-2597
  3. Moström P, Ahlén E, Imberg H, Hansson PO, Lind M. Adherence of self-monitoring of blood glucose in persons with type 1 diabetes in Sweden. BMJ Open Diabetes Res Care 2017;5(1):e000342. https://doi.org/10.1136/bmjdrc-2016-000342
  4. Vincze G, Barner JC, Lopez D. Factors associated with adherence to self-monitoring of blood glucose among persons with diabetes. Diabetes Educ 2004;30(1):112–25. https://doi.org/10.1177/ 014572170403000119
  5. Hirsch I. Introduction: history of glucose monitoring. ADA Clinical Compendia 2018 Aug;1–1. https://doi.org/10.2337/db20181-1
  6. Leelarathna L, Wilmot EG. Flash forward: a review of flash glucose monitoring. Diabetic Medicine 2018;35(4):472–82. https://doi.org/10.1111/dme.13584
  7. Thabit H, Bally L, Hovorka R. Available at a flash: a new way to check glucose. Lancet 2016;388(10057):2213–4. https://doi.org/10.1016/S0140-6736(16)31582-3
  8. freestyle libre 2 can now work as a real-time continuous glucose monitor. accessed on 2300 . 2023. https://www.diabetes.org.uk/about_us/news/freestyle-libre-2-can-now-work-real-time-continuous-glucose-monitor#:~:text=On%203%20August%2C%20Abbott%20announced,FreeStyle%20Libre%202%20sensor%20users.
  9. Reiterer F, Polterauer P, Schoemaker M, et al. Significance and reliability of MARD for the accuracy of CGM systems. J Diabetes Sci Technol 2017;11(1):59–67. https://doi.org/10.1177/1932296816662047
  10. Kovatchev BP, Patek SD, Ortiz EA, Breton MD. Assessing sensor accuracy for non-adjunct use of continuous glucose monitoring. Diabetes Technol Ther 2015;17(3):177-86. https://doi.org/10.1089/dia.2014.0272
  11. Obermaier K, Schmelzeisen-Redeker G, Schoemaker M, et al. Performance evaluations of continuous glucose monitoring systems: precision absolute relative deviation is part of the assessment. J Diabetes Sci Technol 2013;7(4):824–32. https://doi.org/10.1177/193229681300700404
  12. Diabetes Nurse Forum CGM comparison Chart Version 5, 2023. https://www.diabetesspecialistnurseforumuk.co.uk/whats-new
  13. The Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 2008;359(14):1464–76. https://doi.org/10.1056/NEJMoa0805017
  14. Lind M, Polonsky W, Hirsch IB, et al. Continuous glucose monitoring vs conventional therapy for glycaemic control in adults with type 1 diabetes treated with multiple daily insulin injections. JAMA 2017; 317(4):379. https://doi.org/10.1001/jama.201619976
  15. Beck RW, Riddlesworth T, Ruedy K, et al. Effect of continuous glucose monitoring on glycaemic control in adults with type 1 diabetes using insulin injections. JAMA 2017;317(4):371-8. https://doi.org/10.1001/jama.2016.19975
  16. Pratley RE, Kanapka LG, Rickels MR, et al. Effect of continuous glucose monitoring on hypoglycemia in older adults with type 1 diabetes. JAMA 2020;323(23):2397-2406. https://doi.org/10.1001/jama.2020.6928
  17. Laffel LM, Kanapka LG, Beck RW, et al. Effect of continuous glucose monitoring on glycaemic control in adolescents and young adults with type 1 diabetes. JAMA 2020;323(23):2388-96. https://doi.org/10.1001/jama.2020.6940
  18. Thabit H, Prabhu JN, Mubita W, et al. Use of factory-calibrated real- time continuous glucose monitoring improves time in target and HbA1c in a multiethnic cohort of adolescents and young adults with type 1 diabetes: the MILLENNIALS study. Diabetes Care 2020; 43(10):2537–43. https://doi.org/10.2337/dc20-0736
  19. Phillip M, Nimri R, Bergenstal RM, et al. Consensus recommendations for the use of automated insulin delivery technologies in clinical practice. Endocr Rev 2023;44(2):254–80. https://doi.org/10.1210/endrev/bnac022
  20. Oskarsson P, Antuna R, Geelhoed-Duijvestijn P, Krӧger J, Weitgasser R, Bolinder J. Impact of flash glucose monitoring on hypoglycaemia in adults with type 1 diabetes managed with multiple daily injection therapy: a pre-specified subgroup analysis of the IMPACT randomised controlled trial. Diabetologia 2018;61(3):539–50. https://doi.org/10.1007/s00125-017-4527-5
  21. Elliott RA, Rogers G, Evans ML, et al. Estimating the cost‐effectiveness of intermittently scanned continuous glucose monitoring in adults with type 1 diabetes in England. Diabetic Medicine 2024;41(3):e15232. https://doi.org/10.1111/dme.15232
  22. Leelarathna L, Evans ML, Neupane S, et al. Intermittently scanned continuous glucose monitoring for type 1 diabetes. N Engl J Med 2022;387(16):1477–87. https://doi.org/10.1056/NEJMoa2205650
  23. Visser MM, Charleer S, Fieuws S, et al. Effect of switching from intermittently scanned to real-time continuous glucose monitoring in adults with type 1 diabetes: 24-month results from the randomised ALERTT1 trial. Lancet Diabetes Endocrinol 2023; 11(2):96–108. https://doi.org/10.1016/S2213-8587(22)00352-7
  24. Shah N, Deshmukh H, Wilmot EG, et al. The long‐term impact of glucose monitoring with the FreeStyle Libre on glycaemic control and hypoglycaemia awareness in people with type 1 diabetes: insights from the Association of British Clinical Diabetologists national audit. Diabetic Medicine 2023;40(6):e15070. https://doi.org/10.1111/dme.15070
  25. Charleer S, Mathieu C, Nobels F, et al. Effect of continuous glucose monitoring on glycaemic control, acute admissions, and quality of life: a real-world study. J Clin Endocrinol Metab 2018;103(3):1224–32. https://doi.org/10.1210/jc.2017-02498
  26. Leelarathna L, Choudhary P, Wilmot EG, et al. Hybrid closed‐loop therapy: where are we in 2021? Diabetes Obes Metab 2021; 23(3):655–60. https://doi.org/10.1111/dom.14273
  27. Diabetes Specialist Nurse Forum UK. Version 5.0. Diabetes Specialist Nurse Forum comparison chart of Hybrid Closed Loop systems. https://www.diabetesspecialistnurseforumuk.co.uk/whats-new
  28. Braune K, Lal RA, Petruželková L, et al. Open-source automated insulin delivery: international consensus statement and practical guidance for health-care professionals. Lancet Diabetes Endocrinol 2022;10(1):58–74. https://doi.org/10.1016/S2213-8587(21)00267-9
  29. Burnside MJ, Lewis DM, Crocket HR, et al. Open-source automated insulin delivery in type 1 diabetes. N Engl J Med 2022;387(10):869–81. https://doi.org/10.1056/NEJMoa2203913
  30. Ware J, Boughton CK, Allen JM, et al. Cambridge hybrid closed-loop algorithm in children and adolescents with type 1 diabetes: a multicentre 6-month randomised controlled trial. Lancet Digit Health 2022;4(4):e245–55. https://doi.org/10.1016/S2589-7500(22)00020-6
  31. Stewart ZA, Wilinska ME, Hartnell S, et al. Closed-loop insulin delivery during pregnancy in women with type 1 diabetes. N Engl J Med 2016;375(7):644–54. https://doi.org/10.1056/NEJMoa1602494
  32. Lee TTM, Collett C, Bergford S, et al. Automated insulin delivery in women with pregnancy complicated by type 1 diabetes. N Engl J Med 2023;389(17):1566–78. https://doi.org/10.1056/NEJMoa2303911
  33. Tauschmann M, Thabit H, Bally L, et al. Closed-loop insulin delivery in suboptimally controlled type 1 diabetes: a multicentre, 12-week randomised trial. Lancet 2018;392(10155):1321–9. https://doi.org/10.1016/S01406736(18)31947-0
  34. Bergenstal RM, Nimri R, Beck RW, et al. A comparison of two hybrid closed-loop systems in adolescents and young adults with type 1 diabetes (FLAIR): a multicentre, randomised, crossover trial. Lancet 2021; 397(10270):208–19. https://doi.org/10.1016/S0140-6736(20)32514-9
  35. Choudhary P, Kolassa R, Keuthage W, et al. Advanced hybrid closed loop therapy versus conventional treatment in adults with type 1 diabetes (ADAPT): a randomised controlled study. Lancet Diabetes Endocrinol 2022;10(10):720–31. https://doi.org/10.1016/S2213-8587(22)00212-1
  36. Brown SA, Kovatchev BP, Raghinaru D, et al. Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med 2019;381(18):1707–17. https://doi.org/10.1056/NEJMoa1907863
  37. Breton MD, Kovatchev BP. One year real-world use of the Control- IQ advanced hybrid closed-loop technology. Diabetes Technol Ther 2021;23(9):601–8. https://doi.org/10.1089/dia.20210097
  38. Brown SA, Forlenza GP, Bode BW, et al. Multicenter trial of a tubeless, on-body automated insulin delivery system with customizable glycaemic targets in pediatric and adult participants with type 1 diabetes. Diabetes Care 2021;44(7):1630–40. https://doi.org/10.2337/dc21-0172
  39. Crabtree TSJ, Griffin TP, Yap YW, et al. Hybrid closed-loop therapy in adults with type 1 diabetes and above-target HbA1c: a real-world observational study. Diabetes Care 2023;46(10):1831–8. https://doi.org/10.2337/dc23-0635
  40. NHS runs world-first test into ‘sci-fi like’ artificial pancreases, 2022.
  41. Sy SL, Munshi MM, Toschi E. Can smart pens help improve diabetes management? J Diabetes Sci Technol 2020;16(3):628–34. https://doi.org/10.1177/1932296820965600
  42. Munshi MN, Slyne C, Greenberg JM, et al. Nonadherence to insulin therapy detected by Bluetooth-enabled pen cap Is associated with poor glycaemic control. Diabetes Care 2019;42(6):1129–31. https://doi.org/10.2337/dc18-1631
  43. Gomez-Peralta F, Abreu C, Fernández-Rubio E, et al. Efficacy of a connected insulin pen cap in people with noncontrolled type 1 diabetes: a multicenter randomized clinical trial. Diabetes Care 2023;46(1):206–08. https://doi.org/10.2337/dc22-0525
  44. Speight J, Choudhary P, Wilmot EG, et al. Impact of glycaemic technologies on quality of life and related outcomes in adults with type 1 diabetes: a narrative review. Diabetic Medicine 2023; 40(1):e14944. https://doi.org/10.1111/dme.14944
  45. https://www.diabeteseducator.org/danatech/glucose-monitoring/continuous-glucose-monitors-(cgm)/cgm-101/pros-cons-of-cgm#:~:text=Potential%20DISADVANTAGES. accessed on 10/11/2023 at 1400. WEIGHING CGM PROS AND CONS.
  46. Andrew E, Crabtree TSJ, Richardson P, et al. Poster abstracts: Retinopathy outcomes associated with hybrid-closed loop therapy in the NHS England pilot at the University Hospitals of Derby and Burton NHS Trust (UHDB): single centre report from the Association of British Clinical Diabetologists (ABCD) audit. Diabetic Medicine 2023;40(S1):P256. https://doi.org/1111/dme.15048
  47. Type 1 diabetes in adults: diagnosis and management NICE guideline [NG17]. Published 26 August 2015. Last updated: 17 August 2022. Continuous glucose monitoring. https://www.nice.org.uk/guidance/ng17/chapter/Recommendations#blood-glucose-management.
  48. NICE Guidelines. Technology appraisal guidance [TA943]. Last seen at 17/01/2024 @ 1720. 2023. Hybrid closed loop systems for managing blood glucose levels in type 1 diabetes. https://doi.org/10.2337/dc19-2267
  49. Wilson LM, Jacobs PG, Ram Esey KL, et al. Dual-hormone closed- loop system using a liquid stable glucagon formulation versus insulin-only closed-loop system compared with a predictive low glucose suspend system: an open-label, outpatient, single-center, crossover, randomized controlled trial. Diabetes Care 2020; 43(11):2721–9. https://doi.org/10.237/dc19-2267
  50. Castellanos LE, Balliro CA, Sherwood JS, et al. Performance of the insulin-only iLet bionic pancreas and the bihormonal iLet using dasiglucagon in adults with type 1 diabetes in a home-use setting. Diabetes Care 2021;44(6):e118–20. https://doi.org/10.2337/dc20-1086
  51. Xiao Z, Tan X, Chen X, et al. An implantable RFID sensor tag toward continuous glucose monitoring. IEEE J Biomed Health Informatics 2015;19:910–19. https://doi.org/10.1109/JBHI.2015.2415836